Devices/Recall·openFDA/device

Abiomed, Inc. — Class I: ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 1

HighPublished Jul 22, 2026· AI-analyzed Jul 27, 2026View original FDA source
AI-generated regulatory interpretation. The four sections below are an analyst-style summary produced by an AI model from the original FDA source. Always verify against the source before any regulatory, clinical, or commercial decision.
What happened

Abiomed, Inc. has initiated a Class I recall for the Impella CP Introducer Kit 14 Fr due to a potential for sheath leakage under the cap and along hub score lines. The issue affects both 14Fr and 23Fr introducers, which are included in every pump set and possess manufacturing defects that are not visually detectable by the user.

Who it affects

This recall affects Abiomed, Inc., clinical staff utilizing Impella CP heart pump systems, hospital inventory managers, and regulatory/quality departments responsible for mechanical circulatory support devices.

Why it matters

A Class I designation indicates the FDA's assessment of a high risk for serious adverse health consequences or death. The fact that the defect is not visually detectable suggests a significant challenge for point-of-use mitigation, likely requiring a full removal or replacement of affected lots. Regulatory teams may need to investigate manufacturing process controls and supplier quality related to the hub score lines and sheath cap assembly.

Practical takeaway

QA teams should immediately verify inventory for the specified pump set codes and assess the adequacy of current visual inspection protocols, as the source indicates these defects are not visually detectable by users. Regulatory departments should update risk files to account for potential sheath leakage and ensure documentation meets Class I recall reporting requirements.

FDA source material

Class I. Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.. Product: ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories. Firm: Abiomed, Inc. (Danvers MA United States). Status: Ongoing.

Open in openFDA / FDA.gov
AI-generated interpretation. Always verify critical decisions against the original FDA source. Generated with google/gemini-3-flash-preview.